Accuracy of color Doppler velocity in the flow field proximal to a regurgitant orifice: Implications for color Doppler quantitation of valvular incompetence



Accuracy of color Doppler velocity in the flow field proximal to a regurgitant orifice: Implications for color Doppler quantitation of valvular incompetence



Ultrasound in Medicine and Biology 22(5): 605-621



Color Doppler is routinely used in estimates of valvular regurgitation. Velocity and subsequently flow measurements are made at about 7-10 cm from the ultrasonic transducer. Error in velocity measurement may occur due to spatial broadening of the color Doppler beam in the axial, azimuthal and lateral directions. Error in velocity may also occur due to wall filters since the filtering process is not uniform throughout the velocity range indicated by the color bar. An attempt to estimate this error was made using an in vitro orifice model, a numerical finite element model (FEM), and information from the manufacturer. We found that the acoustic beam spatial expansion, wall filter sensitivity and Nyquist limit (NYL) have to be considered simultaneously to account for errors. The combined spatial expansion and wall filter effect on velocity was estimated as a weighted average over the sample volume. The error distributions are not universal but depend on orifice size and flow. For a 3-mm orifice and 100 cm s NYL the overall effect was overestimation of low velocities and significant underestimation of high velocities due to the high velocity gradients inside the sample volume. For the 5- and the 10-mm orifice the effect was less accentuated. Based on this overall error distribution, a correction was incorporated on color Doppler obtained data. The incorporated correction yielded better agreement with numerical velocity data. This correction is important in the application of the proximal isovelocity surface area (PISA) technique and the evaluation of regurgitant flowrates.

Please choose payment method:






(PDF emailed within 0-6 h: $19.90)

Accession: 008095375

Download citation: RISBibTeXText

PMID: 8865557

DOI: 10.1016/0301-5629(96)00029-4


Related references

Color flow doppler quantitation of regurgitant flow rate using the flow convergence region proximal to the orifice of a regurgitant jet. Circulation 78(4 PART 2): II609, 1988

Limitations of color flow Doppler imaging in the quantification of valvular regurgitation: velocity of regurgitant jet, rather than volume, determines size of color Doppler image. American Heart Journal 126(1): 168-176, 1993

Application of color doppler flow imaging to calculate effective regurgitant orifice area an assessment of valvular incompetence independent of hemodynamics. Journal of the American College of Cardiology 19(3 SUPPL A): 297A, 1992

Color Doppler velocity accuracy proximal to regurgitant orifices: Influence of orifice aspect ratio. Ultrasound in Medicine & Biology 25(5): 771-792, 1999

Color flow doppler quantitation of mitral regurgitation using an acceleration flow proximal to the orifice of a regurgitant jet. Circulation 80(4 SUPPL 2): II577, 1989

A new method for quantitation of mitral regurgitation based on color flow Doppler imaging of flow convergence proximal to regurgitant orifice. Circulation 84(4): 1481-1489, 1991

Laser doppler color flow mapping of models of regurgitant jets implications for ultrasound color doppler. European Heart Journal 13(ABSTR SUPPL): 445, 1992

Color encoding of in vitro regurgitant jets by color doppler flow mapping an index of velocity flow rate or regurgitant volume?. Journal Of The American College Of Cardiologysuppl. A: 97a, 1988

Determination of flow rate and orifice area in mitral stenosis by Doppler color flow imaging Quantitative application of centerline velocity/distance function of proximal flow field. Journal of the American College of Cardiology 0(SPEC ISSUE): 379A-380A, 1995

Dynamic changes in mitral regurgitant orifice area A clinical study using color Doppler proximal flow convergence method. Circulation 86(4 SUPPL 1): I257, 1992

A new method for quantification of regurgitant flow rate using color doppler flow imaging of the flow convergence region proximal to a discrete orifice a in vitro study. Circulation 83(2): 594-604, 1991

A new method for quantification of regurgitant flow rate using color Doppler flow imaging of the flow convergence region proximal to a discrete orifice. An in vitro study. Circulation 83(2): 594-604, 1991

Aortic regurgitant flow by color Doppler measurement of the local velocity 7 mm above the leak orifice: Part 1: In vitro measurements. Zeitschrift fuer Kardiologie 88(9): 631-639, Sept, 1999

Color Doppler diagnosis of mechanical prosthetic mitral regurgitation: usefulness of the flow convergence region proximal to the regurgitant orifice. American Heart Journal 120(5): 1137-1142, 1990

Aortic regurgitant flow by color Doppler measurement of the local velocity 7 mm above the leak orifice: Part 2: Comparison with cardiac catheterization. Zeitschrift fuer Kardiologie 88(11): 896-905, 1999