Determination of the renin inhibitor Ro 42-5892 in human plasma by automated pre-column derivatization, reversed-phase high-performance liquid chromatographic separation and electrochemical detection after post-column irradiation

Leube, J.; Fischer, G.

Journal of Chromatography. B, Biomedical Applications 665(2): 373-381

1995


ISSN/ISBN: 1572-6495
PMID: 7795817
DOI: 10.1016/0378-4347(94)00542-d
Accession: 008448856

Download citation:  
Text
  |  
BibTeX
  |  
RIS

Article/Abstract emailed within 0-6 h
Payments are secure & encrypted
Powered by Stripe
Powered by PayPal

Abstract
The renin inhibitor Ro 42-5892 has been found to be very potent, thereby necessitating a sensitive assay method for the evaluation of its pharmacokinetics in man. We report here the development of a very sensitive and selective HPLC assay for the analysis of this compound in human plasma. Ro 42-5892 was extracted from plasma with dichloromethane, derivatized with 2,4-dinitrofluorobenzene and then chromatographed on a Novapak C-18 column (150 times 3.9 mm I.D.) with acetic acid buffer (pH 7)-acetonitrile (100:85). Detection was performed by irradiation at 254 nm, followed by electrochemical oxidation at 550 mV. The extraction recovery of Ro 42-5892 from human plasma (mean 102%) was quantitative. With this method a limit of quantitation of 0.3 ng/ml was achieved. The assay was linear up to 5 ng/ml, had acceptable inter-assay precision (12.2%) and accuracy (9.3%) and was successfully tested for selectivity. This assay was successfully applied to over 250 samples from a pharmacokinetic study in hypertensive patients.