An accurate calibration method of the multileaf collimator valid for conformal and intensity modulated radiation treatments



An accurate calibration method of the multileaf collimator valid for conformal and intensity modulated radiation treatments



Physics in Medicine and Biology 49(12): 2631-2643



Because for IMRT treatments the required accuracy on leaf positioning is high, conventional calibration methods may not be appropriate. The aim of this study was to develop the tools for an accurate MLC calibration valid for conventional and IMRT treatments and to investigate the stability of the MLC. A strip test consisting of nine adjacent segments 2 cm wide, separated by 1 mm and exposed on Kodak X-Omat V films at Dmax depth, was used for detecting leaf-positioning errors. Dose profiles along the leaf-axis were taken for each leaf-pair. We measured the dose variation on each abutment to quantify the relative positioning error (RPE) and the absolute position of the abutment to quantify the absolute positioning error (APE). The accuracy of determining the APE and RPE was 0.15 and 0.04 mm, respectively. Using the RPE and the APE the MLC calibration parameters were calculated in order to obtain a flat profile on the abutment at the correct position. A conventionally calibrated Elekta MLC was re-calibrated using the strip test. The stability of the MLC and leaf-positioning reproducibility was investigated exposing films with 25 adjacent segments 1 cm wide during three months and measuring the standard deviation of the RPE values. A maximum shift over the three months of 0.27 mm was observed and the standard deviation of the RPE values was 0.11 mm.

(PDF emailed within 0-6 h: $19.90)

Accession: 011746799

Download citation: RISBibTeXText

PMID: 15272678

DOI: 10.1088/0031-9155/49/12/011



Related references

Comparison of the helical tomotherapy against the multileaf collimator-based intensity-modulated radiotherapy and 3D conformal radiation modalities in lung cancer radiotherapy. British Journal of Radiology 84(998): 161-172, 2011

A modified method of planning and delivery for dynamic multileaf collimator intensity-modulated radiation therapy. International Journal of Radiation Oncology Biology Physics 47(1): 241-245, April 1, 2000

The effect of gantry and collimator angles on leaf limited velocity and position in dynamic multileaf collimator intensity-modulated radiation therapy. Physics in Medicine and Biology 55(11): 3101-3113, 2010

Compact multileaf collimator for conformal and intensity modulated fast neutron therapy: electromechanical design and validation. Medical Physics 33(9): 3313-3320, 2006

A deliverable four-dimensional intensity-modulated radiation therapy-planning method for dynamic multileaf collimator tumor tracking delivery. International Journal of Radiation Oncology, Biology, Physics 71(5): 1526-1536, 2008

Clinical delivery of intensity modulated conformal radiotherapy for relapsed or second-primary head and neck cancer using a multileaf collimator with dynamic control. Radiotherapy & Oncology 50(3): 301-314, March, 1999

Characterization of a commercial multileaf collimator used for intensity modulated radiation therapy. Medical Physics 28(5): 752-756, 2001

Dynamic multileaf collimator performance for intensity modulated radiation therapy. International Journal of Radiation Oncology Biology Physics 48(3 Supplement): 341, 2000

Influence of high-definition multileaf collimator for three-dimensional conformal radiotherapy and intensity-modulated radiotherapy of prostate cancer. Nihon Hoshasen Gijutsu Gakkai Zasshi 68(7): 825-834, 2013

Integrated high definition intensity multileaf collimator system which provides improved conformal radiation therapy while minimizing leakage. Official Gazette of the United States Patent & Trademark Office Patents 1254(1), Jan 1, 2002

Impact of multileaf collimator leaf positioning accuracy on intensity modulated radiation therapy. Nihon Hoshasen Gijutsu Gakkai Zasshi 67(5): 497-506, 2011

Minimizing the number of segments in a delivery sequence for intensity-modulated radiation therapy with a multileaf collimator. Medical Physics 28(10): 2113-2120, 2001

Lateral loss and dose discrepancies of multileaf collimator segments in intensity modulated radiation therapy. Medical Physics 30(11): 2959-2968, 2003

Dosimetric and qualitative analysis of kinetic properties of millennium 80 multileaf collimator system for dynamic intensity modulated radiotherapy treatments. Journal of Cancer Research and Therapeutics 3(1): 23-28, 2007

Configuration options for intensity-modulated radiation therapy using multiple static fields shaped by a multileaf collimator. Physics in Medicine and Biology 43(2): 241-260, 1998