Simultaneous determination of nine trace mono- and di-chlorophenols in water by ion chromatography atmospheric pressure chemical ionization mass spectrometry

Jin, M.; Yang, Y.

Analytica Chimica Acta 566(2): 193-199

2006


ISSN/ISBN: 0003-2670
DOI: 10.1016/j.aca.2006.03.017
Accession: 012559621

A novel analytical method was proposed for the rapidly simultaneous determination of nine mono-chlorophenols (MCPs) and di-chlorophenols (DCPs) in water samples using eluent generator ion chromatography (IQ coupled with an atmospheric pressure chemical ionization mass Spectrometry (APCI-MS) in the negative mode. The IC separation was carried out on an lonPaco AS11 analytical column (250 mm x 4.0 mm) using gradient KOH containing 15% acetonitrile as organic modifier at a constant flow rate of 1.0 mUmin. The molecular ions m/z [M - H](-) 127 and 161 were selected for the quantification in selected ion monitoring (SIM) mode for MCPs and DCPs, respectively. The average recoveries were between 80.6% and 92.6%. Within-day and day-to-day relative standard deviations were less than 12. 1 % and 13.3%, respectively. The method allowed the nine objective compounds in water samples to be determined at mu g/L levels. It was confirmed that this method could be used in routine analysis. (c) 2006 Elsevier B.V. All rights reserved.