Development of a Novel Echocardiography Ramp Test for Speed Optimization and Diagnosis of Device Thrombosis in Continuous-Flow Left Ventricular Assist Devices The Columbia Ramp Study



Development of a Novel Echocardiography Ramp Test for Speed Optimization and Diagnosis of Device Thrombosis in Continuous-Flow Left Ventricular Assist Devices The Columbia Ramp Study







This study sought to develop a novel approach to optimizing continuous-flow left ventricular assist device (CF-LVAD) function and diagnosing device malfunctions. In CF-LVAD patients, the dynamic interaction of device speed, left and right ventricular decompression, and valve function can be assessed during an echocardiography-monitored speed ramp test. We devised a unique ramp test protocol to be routinely used at the time of discharge for speed optimization and/or if device malfunction was suspected. The patient's left ventricular end-diastolic dimension, frequency of aortic valve opening, valvular insufficiency, blood pressure, and CF-LVAD parameters were recorded in increments of 400 rpm from 8,000 rpm to 12,000 rpm. The results of the speed designations were plotted, and linear function slopes for left ventricular end-diastolic dimension, pulsatility index, and power were calculated. Fifty-two ramp tests for 39 patients were prospectively collected and analyzed. Twenty-eight ramp tests were performed for speed optimization, and speed was changed in 17 (61%) with a mean absolute value adjustment of 424 211 rpm. Seventeen patients had ramp tests performed for suspected device thrombosis, and 10 tests were suspicious for device thrombosis; these patients were then treated with intensified anticoagulation and/or device exchange/emergent transplantation. Device thrombosis was confirmed in 8 of 10 cases at the time of emergent device exchange or transplantation. All patients with device thrombosis, but none of the remaining patients had a left ventricular end-diastolic dimension slope >?0.16. Ramp tests facilitate optimal speed changes and device malfunction detection and may be used to monitor the effects of therapeutic interventions and need for surgical intervention in CF-LVAD patients.

Please choose payment method:






(PDF emailed within 1 workday: $29.90)

Accession: 036604176

Download citation: RISBibTeXText


Related references

Development of a novel echocardiography ramp test for speed optimization and diagnosis of device thrombosis in continuous-flow left ventricular assist devices: the Columbia ramp study. Journal of the American College of Cardiology 60(18): 1764-1775, 2013

Echocardiographic Ramp test for continuous-flow left ventricular assist devices: do loading conditions matter?. Jacc. Heart Failure 3(4): 291-299, 2016

Loading conditions influence reliability of the echocardiographic ramp test in continuous-flow left ventricular assist devices. Journal of Heart and Lung Transplantation 32(11): 1142-1144, 2014

Left Ventricular Decompression During Speed Optimization Ramps in Patients Supported by Continuous-Flow Left Ventricular Assist Devices: Device-Specific Performance Characteristics and Impact on Diagnostic Algorithms. Journal of Cardiac Failure 21(10): 785-791, 2016

Echocardiographic outflow pump ramp test in centrifugal-flow left ventricular assist device. International Journal of Artificial Organs 2017: 0, 2017

Ramp Study Hemodynamics, Functional Capacity, and Outcome in Heart Failure Patients with Continuous-Flow Left Ventricular Assist Devices. Asaio Journal 62(4): 442-446, 2017

Relation between pressure and volume unloading during ramp testing in patients supported with a continuous-flow left ventricular assist device. Asaio Journal 61(3): 307-312, 2016

HeartWare Left Ventricular Assist Device Pump Thrombosis: A Shift Away From Ramp. Jacc. Heart Failure 3(11): 857-859, 2016

Simultaneous ramp right heart catheterization and echocardiography in a ReliantHeart left ventricular assist device. World Journal of Cardiology 9(1): 55-59, 2017

Value of serial echo-guided ramp studies in a patient with suspicion of device thrombosis after left ventricular assist device implantation. Echocardiography 31(1): E5-E9, 2014

Novel Left Heart Catheterization Ramp Protocol to Guide Hemodynamic Optimization in Patients Supported With Left Ventricular Assist Device Therapy. Journal of the American Heart Association 8(4): E010232, 2019

Investigation of the methods of controlling the biventricular assist device with continuous flow pumps. The motor current amplitude of left ventricular assist devices determines the flow rate of right ventricular assist devices (RVAD). Japanese Journal of Artificial Organs 29(1): 51-56, 2000

Use of Hemodynamic Ramp Study to Optimize Continuous-Flow Ventricular Assist Device in a Fontan Patient. Asaio Journal 2018, 2018

Device thrombosis with continuous-flow left ventricular assist devices: is Kounis syndrome involved?. Artificial Organs 38(6): 521-523, 2015

Device thrombosis in continuous-flow left ventricular assist devices: a new manifestation of Kounis syndrome?. Journal of Heart and Lung Transplantation 33(6): 665, 2015