Prevalence of asymptomatic and electrically undetectable intracardiac inside-out abrasion in silicon-coated Riata® and Riata® ST implantable cardioverter-defibrillator leads



Prevalence of asymptomatic and electrically undetectable intracardiac inside-out abrasion in silicon-coated Riata® and Riata® ST implantable cardioverter-defibrillator leads



International Journal of Cardiology 167(1): 254-257



Recently, several cases of symptomatic and/or electrically detectable intracardiac inside-out abrasions in silicon-coated Riata® and Riata® ST leads have been described. However, the prevalence in asymptomatic patients with unremarkable implantable cardioverter defibrillator (ICD) interrogation is unknown. The aim of this study was to determine the prevalence of asymptomatic and electrically undetectable intracardiac inside-out abrasion in silicon-coated Riata® and Riata® ST leads. All 52 patients with an active silicone-coated Riata® and Riata® ST lead followed up in our outpatient clinic were scheduled for a premature ICD interrogation and a biplane chest radiograph. When an intracardiac inside-out abrasion was suspected, this finding was confirmed by fluoroscopy. Mean time since implantation was 71 ± 18 months. An intracardiac inside-out abrasion was confirmed by fluoroscopy in 6 patients (11.5%). Mean time from lead implantation to detection of intracardiac inside-out abrasion was 79 ± 14 months. In all patients with an intracardiac inside-out abrasion, ICD interrogation showed normal and stable electrical parameters. Retrospectively, in 4 of these 6 patients, a coronary angiography performed 25 ± 18 months before diagnosis of intracardiac inside-out abrasion already showed the defect. Despite undetected intracardiac inside-out abrasion, 2 of these 4 patients experienced adequate antitachycardia pacing and ICD-shocks. ICD leads were replaced in all 6 patients. The prevalence of asymptomatic intracardiac inside-out abrasion in silicon-coated Riata® and Riata® ST leads is higher than 10% when assessed by fluoroscopy, and most intracardiac inside-out abrasions are not detectable by ICD interrogation.

(PDF emailed within 0-6 h: $19.90)

Accession: 036915975

Download citation: RISBibTeXText

PMID: 22236515

DOI: 10.1016/j.ijcard.2011.12.076


Related references

Independent multicenter study of Riata and Riata ST implantable cardioverter-defibrillator leads. Heart Rhythm 10(3): 361-365, 2013

Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads. Heart Rhythm 9(8): 1227-1235, 2013

Transvenous extraction of implantable cardioverter-defibrillator leads under advisory--a comparison of Riata, Sprint Fidelis, and non-recalled implantable cardioverter-defibrillator leads. Heart Rhythm 10(10): 1444-1450, 2014

Transvenous extraction of implantable cardioverter defibrillator leads under advisory a a comparison of Riata, Sprint Fidelis, and non-recalled implantable cardioverter defibrillator leads. 2013

Insulation defects in Riata implantable cardioverter-defibrillator leads. Internal Medicine 51(19): 2689-2694, 2013

Prevalence of externalized conductors in Riata and Riata ST silicone leads: results from the prospective, multicenter Riata Lead Evaluation Study. Heart Rhythm 10(12): 1778-1782, 2014

Mechanical and electrical dysfunction of Riata implantable cardioverter-defibrillator leads. Europace 16(12): 1787-1794, 2015

Extreme externalization of riata intracardiac cardioverter defibrillator leads: a new peril of a troublesome lead. Journal of Cardiovascular Electrophysiology 25(2): 216, 2015

Riata implantable cardioverter-defibrillator lead failure: analysis of explanted leads with a unique insulation defect. Heart Rhythm 9(5): 742-749, 2013

Cable externalization at the proximal portion of the superior vena cava coil in Riata implantable cardioverter defibrillator leads. Journal of Cardiology Cases 15(1): 3-5, 2018

Cable externalization and electrical failure of the Riata family of implantable cardioverter-defibrillator leads: A systematic review and meta-analysis. Heart Rhythm 12(6): 1233-1240, 2016

Riata and Riata ST defibrillator leads failure: cable externalization is one problem, but other electrical failures seem more preoccupant. European Heart Journal 34(Suppl 1): P1396-P1396, 2013

Prospective long-term evaluation of Optim-insulated (Riata ST Optim and Durata) implantable cardioverter-defibrillator leads. Heart Rhythm 11(12): 2156-2162, 2015

Long-Term Performance of the Riata/ST Implantable Cardioverter-Defibrillator Lead. American Journal of Cardiology 117(5): 807-812, 2016

Very late perforation by a Riata implantable cardioverter-defibrillator lead-A case report. Journal of Arrhythmia 34(3): 319-321, 2018