Improvements in determination of cardiac output with a Swan-Ganz catheter



Improvements in determination of cardiac output with a Swan-Ganz catheter



Asaio Transactions 36(3): M684-M686



The time constant for heat transfer may affect exact determination of cardiac output with Swan-Ganz catheters. Commercially available Swan-Ganz catheters are provided with thermistors with varying time constants. Current monitoring of cardiac output is not corrected for these time constants, so the conventional method of determining cardiac output using the equation of Stewart-Hamilton produces marked errors. The authors propose a new method of determining cardiac output with Swan-Ganz catheters with varying time constants from thermal dilution curve data based on Newton's cooling law. Values for blood flow rate determined by the new method using a completely stirred tank of original design, mimicking the natural heart and using bovine blood, are almost the same as values observed at varying saline infusion volumes, saline temperatures, and saline infusion times.

(PDF emailed within 1 workday: $29.90)

Accession: 040382649

Download citation: RISBibTeXText

PMID: 2252783


Related references

Improvements in determination of cardiac output measured with swan ganz catheters. Japanese Journal of Artificial Organs 20(2): 521-525, 1991

Comparison between pulse-indicated continuous cardiac output catheter and Swan-Ganz catheter in evaluating blood volume. Zhongguo Wei Zhong Bing Ji Jiu Yi Xue 24(8): 495-496, 2013

Is the cardiac output obtained from a Swan-Ganz catheter always zero?. Journal of Clinical Monitoring and Computing 22(6): 431-433, 2008

Application of the Swan-Ganz catheter and thermodilution cardiac output measurement in cardiac surgery. Kyobu Geka. Japanese Journal of Thoracic Surgery 29(9): 612-617, 1976

An evaluation of thermodilution cardiac output measurement using the Swan-Ganz catheter. Anaesthesia and Intensive Care 9(3): 208-220, 1981

Nursing care in the use of a Swan-Ganz catheter: cardiac output and arterial pressure. Revista Brasileira de Enfermagem 31(2): 193-201, 1978

Swan-Ganz catheter and computer analysis of cardiac output following open heart surgery. Japanese Circulation Journal 39(7): 893-894, 1975

Quick fix of bent prongs of the thermodilution cardiac output connector of a Swan-Ganz catheter. Journal of Clinical Anesthesia 17(5): 407-408, 2005

Cardiac output measurements in off-pump coronary surgery: comparison between NICO and the Swan-Ganz catheter. European Journal of Anaesthesiology 23(10): 848-854, 2006

Erroneous cardiac output measurement due to loop formation of a Swan-Ganz catheter in the pulmonary artery. Journal of Clinical Anesthesia 16(3): 235-236, 2004

Is the Patients Temperature the Cause of the Underestimation of Cardiac Output Using the NICO System Compared with the Swan Ganz Catheter?. Anesthesiology Abstracts of Scientific Papers Annual Meeting ( ): Abstract No A-493, 2002

Re: "Erroneous cardiac output measurement due to loop formation of a swan-ganz catheter in the pulmonary artery". Journal of Clinical Anesthesia 17(2): 151; Author Reply 151-151; Author Reply 151, 2005

Comparison of the Cardiac Output Measured with the NICO Device or the Swan Ganz Catheter in Major Cardiovascular Surgery. Anesthesiology Abstracts of Scientific Papers Annual Meeting ( ): Abstract No A-479, 2002

Monitoring cardiac output with a new hot film anemometer probe mounted on a swan ganz thermo dilution catheter. Artificial Organs 5(2): 208, 1981

Comparison of the Cardiac Output Measured with the NICO Device or the Swan Ganz Catheter in Major Orthopedic Surgery. Anesthesiology Abstracts of Scientific Papers Annual Meeting ( ): Abstract No A-583, 2003