Evaluation of disc retention of Björk-Shiley Delrin heart valves



Evaluation of disc retention of Björk-Shiley Delrin heart valves



Journal of Heart Valve Disease 5 Suppl 2: S243-S245



Impact wear grooves have been observed in the occluder discs of clinically explanted Björk-Shiley Delrin (BSD) heart valves due to repetitive closing impacts against the inlet strut. We examined whether this impact wear could increase the disc-to-strut gap, which could in turn affect the retention of occluder disc. Three BSD heart valves with deep wear grooves were tested in a left ventricular assist device set-up simulating dynamic conditions that would facilitate the dislodgement. The disc was found to be rotating within the flange during the testing and no embolization of the occluder disc occurred in any of the BSD valves tested. The Björk retention limit and the physical retention limit were calculated for each valve. The Björk retention limit is defined as (disc well depth)-(disc-to strut gap)-(1/2 outlet strut wire diameter) and the physical retention as (disc well depth)-(disc-to strut gap). The disc retention would not be affected by the impact wear grooves as long as the disc-to-strut gap did not exceed the physical retention limit.

(PDF emailed within 1 workday: $29.90)

Accession: 046009283

Download citation: RISBibTeXText

PMID: 8905525


Related references

Ex-vivo characterization of three Björk-Shiley Delrin heart valves. Journal of Heart Valve Disease 17(3): 325-331, 2008

Report of a case with re-mitral valve replacement for Björk-Shiley Delrin disc valve prosthesis--disc wear of the Delrin disc 20 years after implantation. Nihon Kyobu Geka Gakkai 44(4): 545-547, 1996

Clinical experience with the Björk-Shiley Delrin tilting disc heart valve. Journal of Heart Valve Disease 5 Suppl 2: S169-S177, 1996

Theory of wear as related to the Björk-Shiley Delrin disc heart valve. Journal of Heart Valve Disease 5 Suppl 2: S216-S228, 1996

Concentric wear of the Delrin disc in a Bjork-Shiley heart valve prosthesis: Report of two cases. Surgery Today (Tokyo) 29(10): 1115-1119, 1999

The Björk-Shiley Delrin tilting disc heart valve: historical perspective, design and need for scientific analyses after 25 years. Journal of Heart Valve Disease 5 Suppl 2: S157-S168, 1996

Late malfunction of the Björk-Shiley valve prosthesis due to Delrin disc defacement. Kyobu Geka. Japanese Journal of Thoracic Surgery 46(12): 1059-1062, 1993

The 37-year durability of a Björk-Shiley Delrin-disc aortic valve prosthesis. Texas Heart Institute Journal 39(2): 284-285, 2012

Concentric wear of the Delrin disc of a Björk-Shiley prosthesis: an uncommon cause of prosthetic incompetence. Journal of Thoracic and Cardiovascular Surgery 107(1): 318-319, 1994

A case report of valve dysfunction associated with abrasion of the Delrin disk used in early Björk-Shiley mitral valves requiring resurgery. Annals of Thoracic and Cardiovascular Surgery 15(2): 126-128, 2009

Complete Transversal Disc Fracture in a Björk-Shiley Delrin Mitral Valve Prosthesis 43 Years After Implantation. Annals of Thoracic Surgery 102(4): E283-E285, 2017

Results 5 to 7 years after aortic valve replacement with the original delrin disc model bjork shiley prosthesis. Scandinavian Journal of Thoracic & Cardiovascular Surgery 11(3): 177-180, 1977

Results five to seven years after aortic valve replacement with the original Delrin disc model Björk-Shiley prosthesis. Scandinavian Journal of Thoracic and Cardiovascular Surgery 11(3): 177-180, 1977

Isolated mitral valve replacement with the Björk-Shiley tilting disc valve prosthesis. A six-year review and a comparison between the Delrin and the pyrolytic carbon disc models. Scandinavian Journal of Thoracic and Cardiovascular Surgery 11(3): 181-185, 1977