Evaluation of an automated seed loader for seed calibration in prostate brachytherapy



Evaluation of an automated seed loader for seed calibration in prostate brachytherapy



Journal of Applied Clinical Medical Physics 7(1): 115-125



Automated seed loaders for permanent prostate implants are now commercially available. Besides improved radiation safety, these systems offer seed assay capability and ease of needle loading, making preplanned as well as intra-operative implant procedures more time-efficient. The Isoloader (Mentor Corp., CA) uses individual I125 seeds (SL-125 ProstaSeed) loaded in up to 199 chambers inside a shielded cartridge. The unit performs seed counting and calibration using a builtin solid-state detector. In order to evaluate the reproducibility and accuracy of the calibration process, two test cartridges were measured with the Isoloader itself and compared with a well-type ionization chamber (HDR-1000Plus, Standard Imaging). The air kerma strength measurements for all seeds using the Isoloader had a standard deviation of about 2.7%. For the eight seeds assayed more intensively using both the Isoloader and well chamber, the standard deviations of the measurements for each seed were in the range of 0.8% to 2.8% and 0.6% to 1.3%, respectively. The variation in the Isoloader calibration is attributed to small detector solid angle and bead geometry within seed capsules (verified by radiographs). The reproducibility of the air kerma strength measured by the Isoloader was comparable to that from the well chamber and was clinically acceptable. Seed strength measured with the Isoloader was on average 1% 2% larger than that measured with the well chamber, indicating that the accuracy of the Isoloader was clinically acceptable.

Please choose payment method:






(PDF emailed within 0-6 h: $19.90)

Accession: 048994385

Download citation: RISBibTeXText

PMID: 16518323

DOI: 10.1120/jacmp.v7i1.2114


Related references

Refining prostate seed brachytherapy: Comparing high-, intermediate-, and low-activity seeds for I-125 permanent seed prostate brachytherapy. BrachyTherapy 14(3): 329-333, 2015

Prostate brachytherapy seed localization with Gaussian blurring and camera self-calibration. Medical Image Computing and Computer-Assisted Intervention 11(Pt 2): 636-643, 2008

Automated treatment planning engine for prostate seed implant brachytherapy. International Journal of Radiation Oncology Biology Physics 43(3): 647-652, 1999

First report on the use of a thinner (125)I radioactive seed within 20-gauge needles for permanent radioactive seed prostate brachytherapy: evaluation of postimplant dosimetry and acute toxicity. BrachyTherapy 12(4): 375-381, 2013

Sequential evaluation of prostate edema after permanent seed prostate brachytherapy using CT-MRI fusion. International Journal of Radiation Oncology Biology Physics 62(4): 974-980, 2005

Selective identification of different brachytherapy sources, ferromagnetic seeds, and fiducials in the prostate using an automated seed sorting algorithm. BrachyTherapy 3(2): 106-112, 2004

Comparison of two treatment approaches for prostate cancer: intensity-modulated radiation therapy combined with 125I seed-implant brachytherapy or 125I seed-implant brachytherapy alone. Journal of Applied Clinical Medical Physics 9(2): 2283, 2008

Permanent prostate brachytherapy (PPB) as it concerns seed migration and extraprostatic seed placement. Cancer Journal 11(5): 432; Author Reply 433, 2005

Dosimetry accuracy as a function of seed localization uncertainty in permanent prostate brachytherapy: increased seed number correlates with less variability in prostate dosimetry. Physics in Medicine and Biology 52(11): 3105-3119, 2007

Is a loose-seed nomogram still valid for prostate brachytherapy in a stranded-seed era?. International Journal of Radiation Oncology Biology Physics 72(2): 623-627, 2008

CT, MR, and ultrasound image artifacts from prostate brachytherapy seed implants: the impact of seed size. Medical Physics 39(4): 2061-2068, 2012

Prostate brachytherapy seed localization by analysis of multiple projections: identifying and addressing the seed overlap problem. Medical Physics 31(5): 1277-1287, 2004

Seed migration after transperineal interstitial prostate brachytherapy by using loose seeds: Japanese prostate cancer outcome study of permanent iodine-125 seed implantation (J-POPS) multi-institutional cohort study. Radiation Oncology 10: 228, 2015

CT- and MRI-based seed localization in postimplant evaluation after prostate brachytherapy. BrachyTherapy 12(6): 580-588, 2014

Automated intraoperative calibration for prostate cancer brachytherapy. Medical Physics 38(11): 6285-6299, 2011