Bias due to Preanalytical Dilution of Rodent Serum for Biochemical Analysis on the Siemens Dimension Xpand Plus



Bias due to Preanalytical Dilution of Rodent Serum for Biochemical Analysis on the Siemens Dimension Xpand Plus



Frontiers in Veterinary Science 5: 3-3



Clinical pathology testing of rodents is often challenging due to insufficient sample volume. One solution in clinical veterinary and exploratory research environments is dilution of samples prior to analysis. However, published information on the impact of preanalytical sample dilution on rodent biochemical data is incomplete. The objective of this study was to evaluate the effects of preanalytical sample dilution on biochemical analysis of mouse and rat serum samples utilizing the Siemens Dimension Xpand Plus. Rats were obtained from end of study research projects. Mice were obtained from sentinel testing programs. For both, whole blood was collected via terminal cardiocentesis into empty tubes and serum was harvested. Biochemical parameters were measured on fresh and thawed frozen samples run straight and at dilution factors 2-10. Dilutions were performed manually, utilizing either ultrapure water or enzyme diluent per manufacturer recommendations. All diluted samples were generated directly from the undiluted sample. Preanalytical dilution caused clinically unacceptable bias in most analytes at dilution factors four and above. Dilution-induced bias in total calcium, creatinine, total bilirubin, and uric acid was considered unacceptable with any degree of dilution, based on the more conservative of two definitions of acceptability. Dilution often caused electrolyte values to fall below assay range precluding evaluation of bias. Dilution-induced bias occurred in most biochemical parameters to varying degrees and may render dilution unacceptable in the exploratory research and clinical veterinary environments. Additionally, differences between results obtained at different dilution factors may confound statistical comparisons in research settings. Comparison of data obtained at a single dilution factor is highly recommended.

(PDF same-day service: $19.90)

Accession: 052189231

Download citation: RISBibTeXText

PMID: 29497614

DOI: 10.3389/fvets.2018.00003



Related references

Investigation of unusual high serum indices for lipemia in clear serum samples on siemens analysers dimension. Biochemia Medica 22(3): 352-362, 2012

Adaptation du dosage plasmatique du methotrexate sur Dimension (R) Xpand. Annales de Biologie Clinique 64(5): 471-477, 2006

Adaptation of methotrexate determination in plasma with Dimension Xpand. Annales de Biologie Clinique 64(5): 471-477, 2006

Falsely elevated tacrolimus concentrations on the Dimension Xpand. Clinical Biochemistry 48(16-17): 1210-1210, 2016

Evaluation Of The Automated Acmia Immunoassay For Cyclosporin A On Dimension?? Xpand?? (Dade Behring). Therapeutic Drug Monitoring 27(2): 257-258, 2005

Analytical evaluation of the Abbott ARCHITECT cyclosporine assay in comparison to LC/MS/MS and Dade Dimension Xpand. Clinical Biochemistry 41(14-15): 1274-1275, 2008

Siemens dimension Vista 1500 Intelligent Lab System and BD Vacutainer Rapid Serum Tube improvements in laboratory turnaround time metrics. Clinical Biochemistry 43(9): 777-0, 2010

Falsely elevated whole blood tacrolimus concentrations due to interference in an affinity column-mediated immunoassay method on Xpand dimension. Therapeutic Drug Monitoring 31(2): 267-268, 2009

Serum thyroxine binding capacity-dependent bias in five free thyroxine immunoassays: Assessment with serum dilution experiments and impact on diagnostic performance. Clinical Biochemistry 34(5): 367-371, July, 2001

Analytical and preanalytical biases in serum proteomic pattern analysis for breast cancer diagnosis. Clinical Chemistry 51(8): 1525-1528, 2005

Chasing rainbows. Color reactions of analyses performed on a Siemens Dimension EXL. Veterinary Clinical Pathology 43(1): 4-5, 2014

Cyclosporine Drug Levels: Comparison of the Architect 1000i with the Siemens Dimension RXL System. Journal of Laboratory Physicians 7(2): 137-137, 2015

Simultaneous measurement of testosterone, androstenedione and dehydroepiandrosterone (DHEA) in serum and plasma using Isotope-Dilution 2-Dimension Ultra High Performance Liquid-Chromatography Tandem Mass Spectrometry (ID-LC-MS/MS). Clinica Chimica Acta; International Journal of Clinical Chemistry 438: 157-159, 2015

Alterations due to dilution and anticoagulant effects in hematologic analysis of rodent blood samples on the Sysmex XT-2000iV. Veterinary Clinical Pathology 45(2): 215-224, 2016

Re-assessing the contribution of serum total cholesterol, blood pressure and cigarette smoking to the aetiology of coronary heart disease: impact of regression dilution bias. European Heart Journal 24(19): 1719-1726, 2003