Comparison of fixation stability of resorbable versus titanium plate and screws in mandibular angle fractures



Comparison of fixation stability of resorbable versus titanium plate and screws in mandibular angle fractures



Journal of Oral and Maxillofacial Surgery 67(8): 1644-1648



The aim of this study was to compare the fixation reliability and stability of titanium and resorbable plates and screws by simulating chewing forces. Mandibular angle fractures in 11 sheep hemimandibles were fixed with 4-hole straight titanium plates and 2.0 x 7-mm titanium screws; in addition, 11 hemimandibles were fixed with 4-hole straight resorbable plates and 2.5 x 6-mm resorbable screws according to the Champy technique. The hemimandibles were mounted with a fixation device in a servohydraulic testing unit for compressive testing. Displacement values under 20, 60, 100, 120, 150, and 200 N; maximum displacements; and maximum forces that the model could resist before breakage were recorded and compared. Significant differences were found between resorbable and titanium plates and screws at all forces (20, 60, 100, 120, 150, and 200 N) (P < .05). We found no statistically significant differences in the breaking force and maximum displacement values (displacement values at the breaking forces) between the groups. The stability of mandibular angle fractures with titanium miniplates under simulated chewing forces was significantly higher than with the resorbable system. Metallic and resorbable fixation systems cannot be used interchangeably to treat mandibular angle fractures under similar loading conditions.

Please choose payment method:






(PDF emailed within 0-6 h: $19.90)

Accession: 052234387

Download citation: RISBibTeXText

PMID: 19615576

DOI: 10.1016/j.joms.2009.03.011


Related references

Poster 079: Comparison of Fixation Stability of Resorbable Versus Titanium Plate and Screws in Mandibular Angle Fractures. Journal of Oral and Maxillofacial Surgery 66(8-Supp-S): 115-116, 2008

Comparison of stability of titanium and absorbable plate and screw fixation for mandibular angle fractures. Oral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontics 106(6): 806-811, 2008

Comparison of resorbable plates and titanium plates for fixation stability of combined mandibular symphysis and angle fractures. Journal of the Korean Association of Oral and Maxillofacial Surgeons 40(6): 285-290, 2014

Stability of skeletal Class III malocclusion after combined maxillary and mandibular procedures: titanium versus resorbable plates and screws for maxillary fixation. Journal of Oral and Maxillofacial Surgery 64(4): 642-651, 2006

Computerized analysis of resorbable polymer plates and screws for the rigid fixation of mandibular angle fractures. Journal of Oral and Maxillofacial Surgery 61(4): 481, 2003

Comparison of Mechanical Stabilization of the Mandibular Angulus Fracture Fixation, With Titanium Plates and Screws, Resorbable Plates and Screws, and Bone Adhesives. Journal of Craniofacial Surgery 29(7): 1780-1787, 2018

Biomechanical analysis of titanium fixation plates and screws in mandibular angle fractures. Nigerian Journal of Clinical Practice 19(3): 386-390, 2016

A Comparative Evaluation of Fixation Techniques in Anterior Mandibular Fractures Using 2 mm Monocortical Titanium Miniplates Versus 24 mm Cortical Titanium Lag Screws. 2012

A Comparative Evaluation of Fixation Techniques in Anterior Mandibular Fractures Using 2.0 mm Monocortical Titanium Miniplates Versus 2.4 mm Cortical Titanium Lag Screws. Journal of Maxillofacial and Oral Surgery 11(4): 442-450, 2012

Skeletal stability after mandibular setback surgery: comparison between the hybrid technique for fixation and the conventional plate fixation using an absorbable plate and screws. Journal of Cranio-Maxillo-Facial Surgery 42(4): 351-355, 2014

Evaluation of 2.0-mm Titanium Three-Dimensional Curved Angle Strut Plate in the Fixation of Mandibular Angle Fractures-A Prospective Clinical and Radiological Analysis. Craniomaxillofacial Trauma and Reconstruction 7(2): 119-125, 2014

Aesthetic and Functional Outcome of Zygomatic Fractures Fixation Comparison With Resorbable Versus Titanium Plates. Annals of Plastic Surgery 76(Suppl. 1): S85, 2016

Comparison of functional outcomes and patient-reported satisfaction between titanium and absorbable plates and screws for fixation of mandibular fractures: A one-year prospective study. Journal of Cranio-Maxillo-Facial Surgery 45(5): 704-709, 2017

Biomechanical comparison of fixation stability using a Lisfranc plate versus transarticular screws. Foot and Ankle Surgery 25(1): 71-78, 2019

Can Resorbable Fixation Screws Replace Titanium Fixation Screws? A Nano-Indentation Study. Journal of Oral and Maxillofacial Surgery 74(7): 1421.E1, 2016