National Consultant in Cardiology Experts' Group Guidelines on dealing with patients implanted with some St. Jude Medical Riata and Riata ST leads



National Consultant in Cardiology Experts' Group Guidelines on dealing with patients implanted with some St. Jude Medical Riata and Riata ST leads



Kardiologia Polska 72(6): 576-582



In December 2010 St. Jude Medical informed about higher incidence of silicone insulation abrasion in implantable cardioverter-defibrillator leads Riata/Riata ST. The manifestation of this phenomenon is the externalisation of conductors outside the body of the lead, which is visible in a fluoroscopy. The abrasion could also involve an insulation under high-voltage coil and in the worst case could result in a short circuit within high voltage part of the system. The incidence of this phenomenon varies from part of to several dozen percent according to published papers and becomes higher in a longer follow-up. The highest probability of malfunction in 8 F single coil and the lowest in 7 F dual-coil leads is observed. For the needs of this guidelines all Riata/Riata ST leads were divided into: functioning, damaged but active (visible externalisation but electrically functioning), malfunctioning. In the last case the lead should be removed and a new one implanted (class of indication I) ,although only implantation of a new lead with abandoning malfunctioning one is allowed and should be considered (IIa). In patients with functioning lead extraction with a new lead implantation may be considered during elective replacement only in high risk patients (IIb). In case of damaged but active lead its extraction with the implantation of a new lead during elective replacement of the device should be considered in high risk population (IIa) and may be considered in other patients (IIb). The final decision related to Riata/Riata ST should be individualised and undertaken in co-operation with the patient after detailed assessment of the risk related to each treatment option.

(PDF emailed within 1 workday: $29.90)

Accession: 054550283

Download citation: RISBibTeXText

PMID: 24961459


Related references

Recommendations of the Working Group of Arrhythmias of the German Society of Cardiology on the approach to patients with Riata® and Riata ST® leads (St. Jude Medical). Nucleus of the Working Group of Arrhythmias of the German Society of Cardiology. Herzschrittmachertherapie and Elektrophysiologie 23(2): 107-115, 2013

Prevalence of externalized conductors in Riata and Riata ST silicone leads: results from the prospective, multicenter Riata Lead Evaluation Study. Heart Rhythm 10(12): 1778-1782, 2014

Long-term Electrical Survival Analysis of Riata and Riata ST Silicone Leads—National VA Experience. 2012

Long-term electrical survival analysis of Riata and Riata ST silicone leads: National Veterans Affairs experience. Heart Rhythm 9(12): 1954-1961, 2013

Performance of the St. Jude Medical Riata leads. Heart Rhythm 6(2): 204-209, 2009

FDA viewpoint: St. Jude Medical's Riata ST leads. Pacing and Clinical Electrophysiology 31(1): 16, 2008

High prevalence of insulation failure with externalized cables in St. Jude Medical Riata family ICD leads: fluoroscopic grading scale and correlation to extracted leads. Heart Rhythm 9(8): 1218-1224, 2013

Independent multicenter study of Riata and Riata ST implantable cardioverter-defibrillator leads. Heart Rhythm 10(3): 361-365, 2013

Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads. Heart Rhythm 9(8): 1227-1235, 2013

Riata and Riata ST defibrillator leads failure: cable externalization is one problem, but other electrical failures seem more preoccupant. European Heart Journal 34(Suppl 1): P1396-P1396, 2013

Independent Multicenter Study of Riata and Riata ST ICD Leads. Heart Rhythm 9(9): 1577-1578, 2012

Prevalence of asymptomatic and electrically undetectable intracardiac inside-out abrasion in silicon-coated Riata® and Riata® ST implantable cardioverter-defibrillator leads. International Journal of Cardiology 167(1): 254-257, 2014

Concerns about the Riata ST (St. Jude Medical) ICD lead. Pacing and Clinical Electrophysiology 31(1): 1-2, 2008

Clinical performance of the St. Jude Medical Riata defibrillation lead in a large patient population. Journal of Cardiovascular Electrophysiology 21(5): 551-556, 2010

Could Externalized St. Jude Medical Riata® Lead Be a Culture Medium of a Polymicrobial Endocarditis? A Clinical Case. Case Reports in Cardiology 2017: 8967234, 2017