Simultaneous determination of creatinine and uric acid in urine by liquid chromatography-tandem mass spectrometry with polarity switching electrospray ionization
Kwon, W.; Kim, J.Y.; Suh, S.; In, M.K.
Forensic Science International 221(1-3): 57-64
2012
ISSN/ISBN: 1872-6283 PMID: 22503624 DOI: 10.1016/j.forsciint.2012.03.025
Accession: 055787758
A simple liquid chromatography-tandem mass spectrometric (LC-MS/MS) method was developed and validated to simultaneously determine creatinine (Cr) and uric acid (UA) levels as a confirmatory method for adulteration or dilution of urine. Centrifuged urine samples (10μL) were diluted with 390μL of distilled water. 30μL of internal standard solution (Cr-d(3), 5μg/mL) and 10μL of acetonitrile were added to 20μL aliquots of diluted urine samples and filtered. The samples (1μL) were introduced into LC-MS/MS with no further pretreatment. Cr and UA were separated on a multi-mode ODS column (Scherzo SM-C18, 75mm×2.0mm I.D., 3μm) and quantified by LC-MS/MS with polarity-switching electrospray ionization. Cr requires the positive-ion mode, whereas the negative-ion mode is required for the analysis of UA. The linear ranges were 1.0-300mg/dL for Cr and 0.5-300mg/dL for UA, with good determination coefficients (R(2)≥0.9988). The intra-day and inter-day precision of the analytes was within 13.0% and 14.4%, respectively. The intra-day and inter-day accuracy was -8.8 to 3.7% and -0.3 to 6.6%, respectively. The lower limits of detection (LLODs) were 0.3mg/dL for Cr and 0.07mg/dL for UA. The applicability of the developed method was examined by analyzing urine samples from suspected drug abusers (n=46).