The effect of a PEEP valve on a Laerdal neonatal self-inflating resuscitation bag



The effect of a PEEP valve on a Laerdal neonatal self-inflating resuscitation bag



Journal of Paediatrics and Child Health 46(1-2): 51-56



Self-inflating bags are used to provide ventilation during neonatal resuscitation. However, they cannot provide positive end expiratory pressure (PEEP) unless a PEEP valve is attached. The ability of Laerdal neonatal self-inflating bags fitted with PEEP valves to reliably deliver PEEP is unclear. The aim of this study was to measure the delivered PEEP at different set PEEP levels and inflation rates. We connected disposable and non-disposable 240 mL Laerdal self-inflating resuscitation bags fitted with PEEP valves to a leak-free test lung. We measured PEEP delivered with the valve set at 5, 7 and 10 cm H(2)O whilst inflating the test lung at rates of 20, 40 and 60 min. Studies were done with 8 L/min of gas flow and with no gas flow. The PEEP delivered was close to the set level immediately after inflation but declined rapidly between inflations. The mean PEEP was higher with faster ventilation rates. When PEEP was set at 7 cm H(2)O, using a non-disposable bag, and an inflation rate of 60/min the mean (SD) PEEP was 5.4 (0.19) cm H(2)O. The PEEP delivered was unrelated to the gas flow into the device. The 240 mL Laerdal self-inflating bag with a PEEP valve delivers PEEP that loses pressure quickly. The level of PEEP delivered is less than that set, particularly at rates below 40/min.

Please choose payment method:






(PDF emailed within 0-6 h: $19.90)

Accession: 056296297

Download citation: RISBibTeXText

PMID: 19943861

DOI: 10.1111/j.1440-1754.2009.01617.x


Related references

Malfunction of a Laerdal resuscitation valve. Anaesthesia 46(6): 505-506, 1991

The self-inflating resuscitation bag delivers high oxygen concentrations when used without a reservoir: implications for neonatal resuscitation. Respiratory Care 54(12): 1665-1670, 2009

Use of self-inflating bags for neonatal resuscitation. Resuscitation 67(1): 109-112, 2005

Oxygen delivery using a neonatal self-inflating resuscitation bag: effect of oxygen flow. Pediatrics 131(4): E1144, 2013

The patient inflating valve in anaesthesia and resuscitation breathing systems. Anaesthesia and Intensive Care 41(2): 163-174, 2013

Failure of an infant Laerdal self-inflating resuscitator due to misassembly. Anaesthesia and Intensive Care 37(3): 500-501, 2009

Oxygen concentration delivery with self-inflating neonatal resuscitation bags. Pediatric Research 35(4 Part 2): 262A, 1994

Oxygen delivery using neonatal self-inflating resuscitation bags without a reservoir. Archives of Disease in Childhood. Fetal and Neonatal Edition 95(5): F315-F319, 2010

Comparison of training in neonatal resuscitation using self inflating bag and T-piece resuscitator. Medical Journal Armed Forces India 71(1): 19-23, 2015

Providing PEEP during neonatal resuscitation: which device is best?. Journal of Paediatrics and Child Health 47(10): 698-703, 2011

Reliability of two common PEEP-generating devices used in neonatal resuscitation. Klinische Padiatrie 221(7): 415-418, 2010

Comparison Of T-Piece Resuscitator With Other Flow - Inflating Bags For Providing Positive Pressure Ventilation During Neonatal Resuscitation: A Systematic Review. Value in Health 18(7): A347, 2015

A modification of the Laerdal infant resuscitator for the simple and safe delivery of CPAP and PEEP. Respiratory Care 29(3): 270-272, 1984

Peritoneal insufflation for laparoscopy with a laerdal resuscitation bag. American Journal of Obstetrics & Gynecology 140(5): 599-601, 1981

Reliability of Single-Use PEEP-Valves Attached to Self-Inflating Bags during Manual Ventilation of Neonates--An In Vitro Study. Plos one 11(2): E0150224, 2016