A Multicenter Performance Evaluation of a Blood Glucose Monitoring System in 21 Leading Hospitals in Spain



A Multicenter Performance Evaluation of a Blood Glucose Monitoring System in 21 Leading Hospitals in Spain



Journal of Diabetes Science and Technology 10(1): 93-100



Self-monitoring of blood glucose is crucial for the effective self-management of diabetes. The present study evaluated the accuracy of the Contour® XT blood glucose monitoring system (BGMS) compared to the reference method in a large multicenter study under routine lab conditions at each hospital site. This study was conducted at 21 leading hospitals in Spain using leftover whole blood samples (n = 2100). Samples were tested with the BGMS using 1 commercial strip lot and the local laboratory hexokinase method. BGMS accuracy was assessed and results were compared to ISO 15197:2013 accuracy limit criteria and by using mean absolute relative difference analysis (MARD), consensus (Parkes) error grid (CEG), and surveillance error grid analyses (SEG). Pooled analysis of 2100 measurements from all sites showed that 99.43% of the BGMS results were within the ranges accepted by the accuracy limit criteria. The overall MARD was 3.85%. MARD was 4.47% for glucose concentrations < 70 mg/dL and 3.81% for concentrations of 70-300 mg/dL. In CEG, most results (99.8%) were within zone A ("no effect on clinical action"); the remaining ones (0.2%) were in zone B ("little to no effect on clinical action"). The SEG analysis showed that most of the results (98.4%) were in the "no risk" zone, with the remaining results in the "slight, lower" risk zone. This is the largest multicenter study of Contour XT BGMS to date, and shows that this BGMS meets the ISO 15197:2013 accuracy limit criteria under local routine conditions in 21 leading Spanish hospitals.

(PDF emailed within 0-6 h: $19.90)

Accession: 057033021

Download citation: RISBibTeXText

PMID: 26253142

DOI: 10.1177/1932296815598777



Related references

Analytical Performance Evaluation of Infopia Element™ Auto-coding Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose. Journal of Clinical Laboratory Analysis 30(6): 849-858, 2016

A Multicenter Evaluation of the Performance and Usability of a Novel Glucose Monitoring System in Chinese Adults With Diabetes. Journal of Diabetes Science and Technology 11(2): 290-295, 2016

Interinstitutional comparison of bedside blood glucose monitoring program characteristics, accuracy performance, and quality control documentation: a College of American Pathologists Q-Probes study of bedside blood glucose monitoring performed in 226 small hospitals. Archives of Pathology & Laboratory Medicine 122(6): 495-502, 1998

Accuracy and user performance evaluation of a blood glucose monitoring system. Diabetes Technology & Therapeutics 17(3): 152-158, 2015

Accuracy and user performance evaluation of the Contour(®) Next Link 2.4 blood glucose monitoring system. Clinica Chimica Acta; International Journal of Clinical Chemistry 448(): 139-145, 2016

Performance evaluation and labeling comprehension of a new blood glucose monitoring system with integrated information management. Journal of Diabetes Science and Technology 5(5): 1144-1153, 2012

Brazilian multicenter study for the evaluation of patients' satisfaction of blood glucose self-monitoring with BGStar(®) blood glucose meter in insulinized patients with diabetes mellitus type 1 and 2. Diabetology & Metabolic Syndrome 8(1): 66-66, 2016

System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197. Journal of Diabetes Science and Technology 6(5): 1060-1075, 2013

Multicenter evaluation of the reliability of five blood glucose monitoring systems. Annales de Biologie Clinique 69(1): 55-61, 2011

Multicenter evaluation of a meter for blood glucose and ketone monitoring. Diabetes 48(SUPPL 1): A62, 1999

Clinical evaluation and alternative site blood glucose testing of the EasyPlus mini R2N blood glucose monitoring system. Clinica Chimica Acta; International Journal of Clinical Chemistry 403(1-2): 167-172, 2009

Multicenter Observational Study of the First-Generation Intravenous Blood Glucose Monitoring System in Hospitalized Patients. Journal of Diabetes Science and Technology 9(4): 739-750, 2016

An Evaluation of Blood Glucose Measurement Using TRUEresult Blood Glucose Monitoring System. Journal of the Medical Association of Thailand 98(9): 839-846, 2015

Evaluation of blood glucose fluctuation in Japanese patients with type 1 diabetes mellitus by self-monitoring of blood glucose and continuous glucose monitoring. Diabetes Research and Clinical Practice 108(2): 342-349, 2015

Effect of ambient temperature on analytical and clinical performance of a blood glucose monitoring system: Omnitest Sensor glucose meter. Annals of Clinical Biochemistry 38(Pt 6): 676-683, 2001