Heimlich valve orientation error leading to radiographic tension pneumothorax: analysis of an error and a call for education, device redesign and regulatory action



Heimlich valve orientation error leading to radiographic tension pneumothorax: analysis of an error and a call for education, device redesign and regulatory action



Emergency Medicine Journal 33(4): 260-267



Medical errors are commonly multifactorial, with adverse clinical consequences often requiring the simultaneous failure of a series of protective layers, termed the Swiss Cheese model. Remedying and preventing future medical errors requires a series of steps, including detection, mitigation of patient harm, disclosure, reporting, root cause analysis, system modification, regulatory action, and engineering and manufacturing reforms. We describe this process applied to two cases of improper orientation of a Heimlich valve in a thoracostomy tube system, resulting in enlargement of an existing pneumothorax and the development of radiographic features of tension pneumothorax. We analyse elements contributing to the occurrence of the error and depict the implementation of reforms within our healthcare system and with regulatory authorities and the manufacturer. We identify features of the Heimlich valve promoting this error and suggest educational, design, and regulatory reforms for enhanced patient safety.

Please choose payment method:






(PDF emailed within 0-6 h: $19.90)

Accession: 057976300

Download citation: RISBibTeXText

PMID: 26531858

DOI: 10.1136/emermed-2015-204821


Related references

Tension pneumothorax after reversal of a Heimlich valve. Ajr. American Journal of Roentgenology 158(4): 763-764, 1992

The management of tension pneumothorax in the neonate using the Heimlich flutter valve. Journal of Pediatrics 84(3): 438-440, 1974

Resolved tension pneumothorax in a newborn with an adult Heimlich valve. Resuscitation 77(3): 294-295, 2008

Iatrogenic tension pneumothorax complicating outpatient Heimlich valve chest drainage. Journal Of Emergency Medicine. 12(4): 477-479, 1994

Malfunction of a Heimlich flutter valve causing tension pneumothorax: case report of a rare complication. Patient Safety in Surgery 4(1): 8, 2010

Error begat error: design error analysis and prevention in social infrastructure projects. Accident; Analysis and Prevention 48: 100-110, 2012

Delayed diagnosis of contralateral tension pneumothorax during thoracotomy--an example of fixation error. Anaesthesia 48(2): 177-178, 1993

Error significance but not error expectancy predicts error-related negativities for different error types. Behavioural Brain Research 297: 259-267, 2016

The role of hand of error and stimulus orientation in the relationship between worry and error-related brain activity: Implications for theory and practice. Psychophysiology 52(10): 1281-1292, 2015

Chemotherapy error: Analysis of an error reporting database, recent case reports, error reduction methods, current reporting systems. Journal of Clinical Oncology 23(16_Suppl): 6091-6091, 2016

Medical error: a discussion of the medical construction of error and suggestions for reforms of medical education to decrease error. Medical Education 35(9): 855-861, 2001

Heimlich valve and pneumothorax. Annals of Translational Medicine 3(4): 54, 2015

Epidural pump programming error leading to inadvertent 10-fold dosing error during epidural labor analgesia with ropivacaine. Journal of Patient Safety 6(4): 244-246, 2010

Epidural Pump Programming Error Leading to Inadvertent 10-Fold Dosing Error During Epidural Labor Analgesia With Ropivacaine. Journal of Patient Safety 6(4): 244-246, 2010

On the approximate identification of the leading coefficient of a linear elliptic equation by output error and equation error minimization. Optimal Control Applications and Methods 10(3): 229-253, 1989