New organ-based tube current modulation method to reduce the radiation dose during computed tomography of the head: evaluation of image quality and radiation dose to the eyes in the phantom study



New organ-based tube current modulation method to reduce the radiation dose during computed tomography of the head: evaluation of image quality and radiation dose to the eyes in the phantom study



La Radiologia Medica 122(8): 601-608



A new organ-based tube current modulation (NOB-TCM) method was designed with the intent to decrease tube current by 30% over a prescribed 90° radial arc across the anterior aspect of the radiosensitive organ, without increasing tube current in the remaining radial arc. We compared a reference scan and five other dose-reducing methods with regard to effects on dose, practicality, and image quality to determine the most effective method for the reduction of the radiation dose to the eyes during CT examinations of the head. We compared the radiation doses to the eyes and physical image quality in different regions of interest for TCM and shielding scans. Three types of TCM scans were performed: longitudinal TCM, angular TCM, and NOB-TCM. A bismuth sheet and lead goggles were each applied for the shielding scan. Relative to the reference scan, the dose to the eye was reduced to 25.88% with NOB-TCM, 44.53% with lead goggles, and 36.91% with a bismuth shield. Relative to the reference scan, the mean signal-to-noise ratio (SNR) was decreased to 8.02% with NOB-TCM, 28.36% with lead goggles, and 32.95% with the bismuth shield. The SNR of the anterior region of interest was decreased to 11.89% with NOB-TCM and 87.89% with the bismuth shield. The average figure of merit was increased by 11.7% with longitudinal TCM and 13.39% with NOB-TCM, compared with the reference scan. NOB-TCM is a superior solution for head CT, including the orbital area, due to the reduction in radiation exposure without significant loss in image quality.

Please choose payment method:






(PDF emailed within 0-6 h: $19.90)

Accession: 060016966

Download citation: RISBibTeXText

PMID: 28341967

DOI: 10.1007/s11547-017-0755-5


Related references

Automatic Tube Current Modulation and Tube Voltage Selection in Pediatric Computed Tomography: A Phantom Study on Radiation Dose and Image Quality. Investigative Radiology 54(5): 265-272, 2019

The application of automatic tube current modulation (ATCM) on image quality and radiation dose at abdominal computed tomography (CT): A phantom study. Journal of X-Ray Science and Technology 21(4): 453-464, 2013

Technical Note: Phantom study to evaluate the dose and image quality effects of a computed tomography organ-based tube current modulation technique. Medical Physics 42(11): 6572-6578, 2015

Reduced radiation exposure of the female breast during low-dose chest CT using organ-based tube current modulation and a bismuth shield: comparison of image quality and radiation dose. Ajr. American Journal of Roentgenology 200(3): 537-544, 2013

The impact of pediatric-specific dose modulation curves on radiation dose and image quality in head computed tomography. Pediatric Radiology 45(12): 1814-1822, 2015

Iterative reconstruction and individualized automatic tube current selection reduce radiation dose while maintaining image quality in 320-multidetector computed tomography coronary angiography. Clinical Radiology 68(11): E570-E577, 2013

Effect of automatic tube current modulation on radiation dose and image quality for low tube voltage multidetector row CT angiography: phantom study. Academic Radiology 16(8): 997, 2009

Radiation exposure and image quality of normal computed tomography brain images acquired with automated and organ-based tube current modulation multiband filtering and iterative reconstruction. Investigative Radiology 47(3): 202-207, 2012

Automated tube voltage adaptation in head and neck computed tomography between 120 and 100 kV: effects on image quality and radiation dose. Neuroradiology 56(9): 797-803, 2014

Reduction of patient radiation dose with a new organ based dose modulation technique for thoraco-abdominopelvic computed tomography (CT) (Liver dose right index). Diagnostic and Interventional Imaging 99(7-8): 483-492, 2018

Comparative Study of Volume Computed Tomography Dose Index and Size-Specific Dose Estimate Head in Computed Tomography Examination for Adult Patients Based on the Mode of Automatic Tube Current Modulation. Medical Science Monitor 25: 71-76, 2019

Knowledge-based iterative model reconstruction: comparative image quality and radiation dose with a pediatric computed tomography phantom. Pediatric Radiology 46(3): 303-315, 2016

Comparative analysis of radiation dose and image quality between organ dose modulation and 3D smart mA modulation during head-neck CT angiography. JournalofX-RayScienceandTechnology27(1):97, 2019

Radiation dose and image quality with abdominal computed tomography with automated dose-optimized tube voltage selection. Journal of International Medical Research 42(4): 1011-1017, 2014

Dose reduction in computed tomography of the chest: image quality of iterative reconstructions at a 50% radiation dose compared to filtered back projection at a 100% radiation dose. Rofo 186(6): 576-584, 2014